ProSens QM-100

ProSens is a new line of modern industrial instruments, which integrates sensors, converters, meters and controllers functionalities. Using the latest miniaturisation technologies these compact instruments are able to be equipped with two independent universal inputs, two binary or two analogue outputs, as well as communication port RS-485 with Modbus protocol. Model QM-100 is a version without probe.

A large built-in display and output signals mean that the ProSens line units find applications in control systems. There are many industrial applications, where ProSens can act as stand-alone controller. It can also cooperate with master devices via Modbus protocol, being part of big network, which makes it perfect instrument for distributed monitoring system

 

Power supply 24V DC (11 ÷ 36V DC); for versions with voltage output min. 16V DC
Type of probe none
Measurement range current: 0/4-20 mA; voltage: 0/1-5 V, 0/2-10V, 0-60 mV, 0-75 mV, 0-100 mV, 0-150 mV; RTD: Pt100, Pt500, Pt1000, measuring range: -100°C ÷ 600°C; thermocouple: type K, S, J, T, N, R, B, E; measuring ranges: -200°C ÷ +1370°C (K); -50°C ÷ +1768°C (S); -210°C ÷ +1200°C (J); -200°C ÷ +400°C (T); -200°C ÷ +1300°C (N); -50°C ÷ +1768°C (R); +250°C ÷ +1820°C (B); -200°C ÷ +1000°C (E)
Power consumption max. 2,5 W
Inputs 1 or 2 universal: current; voltage; RTD; thermocouple K, S, J, T, N, R, B, E types
Binary outputs 0, 1 or 2 electronic NO relays, 24V AC/35V DC, max. 200 mA
Analogue outputs active current: operating range 0/4-20 mA (max. 0-24 mA), load resistance 700 Ω max., resolution 13 bit; passive current: isolated, operating range 4-20 mA (max. 2,8-24 mA), load resistance 600 Ω@24VDC, resolution 13 bit; active voltage: operating range 0/1-5V, 0/2-10V (max. 0-11V), load resistance min. 2000 Ω, resolution 13 bit
Communication interface RS-485, 8N1 and 8N2, 1200 bit/s ÷ 115200 bit/s, Modbus RTU, not galvanically isolated
Operating temperature -30°C ÷ +80°C, case with electronics (out of range -20 ÷ +70°C LCD and IR receiver turn off)
Accuracy accuracy: 0.1% @25°C ± one digit (inputs: current, voltage, milivoltage, thermoresistance, thermocouple K, J, E); 0.2%@ 25°C (thermocouple N), 0.5%@25°C (thermocouple S, T, R, B)
Case material ASA LURAN + polycarbonate
Case wall mounted
Case dimensions 120 x 90 x 50 mm
Protection class IP 67 (version without display); IP 65 (version with display)
Display optionally: graphic LCD, 128 x 64 points, with backlight
Temperature stability typ. 50 ppm/°C
Software free S-Config 2 software for ProSens line can be downloaded directly from “Download” section (top menu)